MAGNET POWER SUPPLY SYSTEM 7000
9
DANFYSIK A/S - DK 4040 JYLLINGE - DENMARK.
2.7
RS422 or RS485 Multidrop configuration
Up to 32 SYSTEM 7000 units can be connected in a RS422 or RS485 multidrop configuration.
It is also possible to run a Master/Slave configuration within a multidrop system. All units
must be connected in order to read the individual unit interlock status signals, but for
controlling the system only the master connection is necessary. Trying to control one of the
slaves will have no effect as the control commands are disabled on the slaves. To set up the
Master/Slave configuration, see chapter 2.6.
•
The Master / Slave slide switch located on the back of the supply must be placed in Master
position for all masters
.
•
Address all units with a unique address.
See SW manual doc- ApBCPsw1c.doc on how to set
up the unit address.
Analog
Parallel units
to slaves
Parallel units
to master
External Interlocks
Digital
AC Main Power
External trig
Factory Program
MPS status
Current
Output
Ground
fuse
Ac control input
Rack Fan
Analog
Parallel units
to slaves
External Interlocks
AC Main Power
External trig
Factory Program
MPS status
Current
Output
Ground
fuse
Ac control input
Rack Fan
Analog
Parallel units
to slaves
External Interlocks
AC Main Power
External trig
Factory Program
MPS status
Current
Output
Ground
fuse
Ac control input
Rack Fan
Parallel units
to master
Parallel units
to master
Master Slave
Master Slave
Analog
Parallel units
to slaves
Parallel units
to master
External Interlocks
Digital
AC Main Power
External trig
Factory Program
MPS status
Current
Output
Ground
fuse
Ac control input
Rack Fan
Master Slave
Digital
Digital
MPS no. 1
addresse : X
MPS no. 2
addresse : X+1
MPS no. 3
addresse : X+2
MPS no. Y
addresse : X+(Y-1)
X :Start address
Y : The last unit of MPS on the
multdrop line
Master Slave
Digital
Example of multidrop configuration
About the termination of RS422 or RS45,
see chapter 3.6.1 “Termination using RS422 or
RS485”