MAGNET POWER SUPPLY SYSTEM 7000
11
DANFYSIK A/S - DK 4040 JYLLINGE - DENMARK.
Mode 1 – ON & OFF through STANDBY
In this mode the power supply must first be set into
the stand by state before entering the ON or OFF
state. An interlock puts the power supply immediately
in the FAULT state.
In the stand by mode is the output converter
without
power.
State
FAN
Output
Converter
OFF OFF Not
Powered
ON ON Powered
STB-1 ON
Not
Powered
FAULT ON
Not
Powered
Mode 2 – ON & OFF through STANDBY
This mode can be seen as a mixture of mode 0 and
1. Here the inserted stand by mode is actually
nearly the same as the ON state but with the output
forced to zero (compared to mode 0). An interlock
puts the power supply immediately in the FAULT
state.
In the stand by mode the output converter is
with
power but forced to zero output.
State
FAN
Output
Converter
OFF OFF Not
Powered
ON ON Powered
STB-2 ON
Powered
FAULT ON
Not
Powered
The STB transition is locally initiated by pressing
the reset button and thereafter at the same time the
ON button. Remotely the STB command must be
issued.
The LED inside the Reset push button indicated that the Power Supply is interlocked either due
to an internal or external fault.
ON-S
STB-1
INTL
{event}
ON
RES
FAULT
RES
{no INTL}
OFF
Else
OFF
OFF
INTL
{event}
AC Power
ON
OFF-S
RES
{no INTL}
Else
Else
INTL
{event}
ON-S
STB-2
INTL
{event}
ON
STB
FAULT
OFF
Else
OFF
AC Power
ON
OFF-S
STB
INTL
{event}
{Out-Dis}
RES
{no INTL}
ON
INTL
{event}
Else
Else
Else