MAGNET POWER SUPPLY SYSTEM 7000
13
DANFYSIK A/S - DK 4040 JYLLINGE - DENMARK.
Gain factor:
→
1
5
10
(factory)
P302 SW1
Closed Closed Open
P302 SW2
Closed Open Open
Frequency Pole
→
Low
(factory)
High
P306 SW1
Closed Open
Current / Voltage mode
This switch setting is for internal use and debugging only by authorized Danfysik service
personnel. In general it sets the unit to a voltage supply where a 20A (full scale) setting gives a
voltage output of approximately 35V.
Loop mode
→
Current mode (factory)
Voltage mode
P305 SW1
Open Closed
P305 SW2
Closed Open
Hint:
The time constant of the magnet (transfer function of the load) can be measured, using
the SYSTEM 7000 in the voltage mode together with an attached gain/phase analyser.
Machine Protection System signal
The Machine Protection System signal (Abbreviated to “M.P.S. signal”) is available through an
Amphenol connector located on the back of the cabinet. This signal is produced by AND’ing;
the three signals; ON status, Sum Interlock and the Main contactor is on.
The M.P.S. signal is high when the SYSTEM 7000 is ON.
Sum interlock and the ON status can be excluded from the “M.P.S. signal”
Sum Interlock/ ON Status Enable / Disable
Switch
↓
Signal
name
↓
Enabled
(factory)
Disabled
P702 SW1
Sum interlock
Closed
Open
P702 SW2
ON status
Closed
Open
Interlock enable – disable - rerouting
Two interlocks (Ground Fault Interlock, Over Voltage) and one status signal (Out of
regulation) can be disabled if not desired.
Interlock/ Status Enable / Disable
Enable
Disable
P801 SW1
Interlock Ground Fault
Open (factory)
Closed
P801 SW2
Status Out of regulation
Open (factory)
Closed
P803-SW2
Interlock Over voltage
Open (factory)
Closed