System 8700 User Manual
Rev 0.95
Page 27 of 32
4.3.2.2.
S3H response
Bit
Interpretation
0
Interlock from regulation board, eIL_RegBoard0
1
Interlock from regulation board, eIL_RegBoard1
2
Interlock from regulation board, eIL_RegBoard2
11
eIL_Interlock2_11
12
eIL_Interlock2_10
13
eIL_Interlock2_09
14
eIL_Interlock2_08
15
eIL_Interlock2_07
16
eIL_Interlock2_06
17
eIL_Interlock2_05
18
eIL_Interlock2_04
19
eIL_Interlock2_03
20
Overload,
1: overload detected
21
Pull up power failure,
1: failure present
22
Soft start failure,
1: failure present
23
Polarity switch failure, 1: failure present
24
PLC module failure,
1: failure present
25
Regulation board interlocks, 1: one or more interlocks present from
regulation board. Use command S3H to read regulation board
interlocks
26
Regulation board comm. failure, 1: failure present
27
Tracking,
1: not tracking
28
Sum interlock,
1: one or more interlocks present
29
Polarity reverse,
1: polarity is reversed
30
Polarity normal,
1: polarity is positive
31
Output On,
1: Output is on