Page 6 of 32
Rev 0.95
DANGER! ELECTRIC SHOCK HAZARD!
Disconnect power at switch board before attempting to work on the Magnet
Power Supply. High voltages are present at the terminals and within the power
supply for up to 10 minutes after disconnection of the electrical supply. Always
ensure by using a suitable measuring device that no voltage is present prior to
commencing work.
CAUTION!
Ensure correct grounding connection. The grounding cable must be selected to
be able to carry the maximum supply fault current. This is normally limited by
fuses at the switch board. Fuses at the switch board must be selected according
to local legislation or code.
About this document
This document is intended for design and service engineers responsible for the installation,
maintenance and repair of the Power Supplies using SYSTEM8700 control system supplied by
Danfysik A/S. The power supply itself and its regulation system is not within the scope of this
document.
1.2.1.
Intended use
The Magnet Power Supply (MPS) is intended for professional incorporation into complete
accelerator systems as a part of a fixed installation. It must be installed in an enclosed
operating area. The power supplies are built according to industrial standards.
The intended use is to control a magnet by supplying the current for the magnet. The MPS
includes an interlock system, which shuts down the power supply in case a number of
predefined errors occur.
Only qualified personnel are allowed to install the MPS, and only instructed personnel are
allowed to operate it.
In case of failure, the accelerator system must have the means to shut down in a safe manner.