MAGNET POWER SUPPLY SYSTEM 9100
11
DANFYSIK A/S - DENMARK.
No other additives should be used.
During operation, the conductivity of the coolant should be checked regularly. A few
days operation with water of poor quality can cause more corrosion damage than during
several years‟ operation with good quality water.
2.4
Installation
Before and during installation of the Power Supply, the following points should be checked /
carried out.
*
Check that the main voltage and frequency matches to the specified and labelled
requirements.
*
Check that screw connections from the output terminals are tightened.
*
Check that all screws on D-sub connectors are tightened.
*
If the magnet is provided with a thermal breaker connect the contacts to external
interlocks and +24 VGND (Ground for +24 V)
*
Unused external interlocks shall be connected to +24V ground (Ground for +24 V pin
12, X112 or to pin 11 +24V_Ext_GND if +24V_Ext, pin X8 is used for interlock
supply)