MAGNET POWER SUPPLY SYSTEM 9100
29
DANFYSIK A/S - DENMARK.
COLD BOOT SETUP
(Esc<COLDBOOT)
After power up or reset the power supply can wake up in different control conditions. The
switch settings shown below describe the possible Wake up conditions.
Parameters:
SW3 FOR COLD BOOT INITIAL SETUP
O
N
1
2
3
4
7
Default setting after “COLD BOOT”
SW4
#1
#1: Local line is allways in auto answer mode.
#2: Using ACK/NACK protocol with or without error code will overide this setting.
#2
REMOTE AUTO ANSWER
DISABLED
¤1: “?”-BELL, Without further information.
¤2: “?”-BELL, Followed by error in code.
¤3: “?”-BELL, Followed by error in text.
ENABLED
¤1
¤2
¤3
¤3
DEFAULT LINE IN
REMOTE
LOCAL
LOCAL ADDRESSING
ENABLED
DISABLED
REMOTE ADDRESSING
ENABLED
DISABLED
Wake to 0%/100%
NOT USED
ERROR RESPONSE
O
N
1
2
3
4
5
6
7
8
#3
0% at power up
100% at power up
#3: Default is 0% output current at power up. [From SW version SCS109]
Remote addressing:
Disabled
Local addressing:
Disabled
Default line in:
Local
Remote Auto answer:
Disabled
Error response:
“?”-BELL only
Wake up output current:
0%