MAGNET POWER SUPPLY SYSTEM 9100
32
DANFYSIK A/S - DENMARK.
AD AUTO ADJUSTMENT (Gain & Offset)
(Esc<ADSET)
The scaling factors for the AD channels can automatically be adjusted with this setting.
SW3 FOR ADSET
O
N
1
2
3
4
14
Parameters
SW4
0
1
30
2
3
31
#1
#1: Not all channels are implemented.
CH_No. 2
3
CH_No. 2
2
CH_No. 2
1
CH_No. 2
0
NOT USED
FUNC
MODE
CH_No. 2
4
O
N
1
2
3
4
5
6
7
8
OFFSET
GAIN
USE
RESTORE
OFFSET adjustment
The OFFSET adjustment can automatically be executed if the value lies between 0 and 255 of
the AD channel resolution.
To activate the automatic OFFSET adjustment following steps must be preformed:
-
Select SW3 as above. “setting number14"
-
Select the desired channel number “lever 1 to 5" on SW4
-
Set lever 8 on SW4 to OFF (The LEDs will display the present Offset value)
-
Ensure that the AD input signal is set to zero. Grounded or turned OFF in any way.
-
Press the Setup Button
-
To restore the Offset value to default factory setting (Zero), please set lever 7 to one
before pressing the Setup Button.
CHANNEL VALUE
CHANNEL
VALUE
0
Output current
1
Field input
2
Output Voltage
3
In15V sup.
4
Internal -15V sup.
5
In5V sup.
6
Delta Temperature
7
I set value
8
Optional Iout (16 Bit)
9
Iout (16 Bit)(Crtl panel).
10
Not used
11
Output current for display
12
Output voltage for display
13
Not used
14
Not used
15
Not used
16
In5Vana.
17 I set value
18
Spare