13

O

2

 sensor temperature sensitivity

Due to the sensor design the readout will be 
affected when moving the device from cold to 
warm temperatures or vice versa.
Normally for compensation to work properly the 
device must be stabilized at the current ambient 
temperature for some time (up to 1-2 hours).
However if you cannot wait for this time, use 
short stabilizing time, for example 10-15 min., 
and perform an offset (20.9%) calibration of the 
O

2

 sensor right before each measurement series 

until device has stabilized.
This will bring the calibration back in line with 
current response.

Calibration

CheckPoint O

2

 only has to be calibrated with 

atmospheric air. If CheckPoint does not show 
approx. 20.9 % O

2

 (between 21.3 % and 20.5 %) 

when measuring atmospheric air, it must be 
calibrated with atmospheric air. Make sure that 
only atmospheric air is sucked during calibration.

Start calibration 

 (2) + 

 (3) :

Press the 

 key and hold it down. Press 

 and 

release both keys. During calibration the display 
alternately shows “CAL.” and “20.9” for 45 
seconds. CheckPoint has now been calibrated.

Stop calibration

Press 

 to interrupt the calibration process.

External power supply/charger (14) 

CheckPoint can be connected to an external 
power supply/charger. When the power supply is 
connected/disconnected, the display shows 
"P.On" and "P.OF.", respectively. If CheckPoint is 
fitted with ordinary batteries, they will be 
disconnected. If CheckPoint is fitted with 
rechargeable batteries, they will be recharged 
automatically.

WARNING! Only connect the original 
power supply.

CheckPoint messages

"P.On "

Power on. External power supply
connected.

"P.OF. "

Power off. External power supply
disconnected.

"Lo./BAt."

Low battery voltage. Replace the
batteries, or connect the external
power supply/charger.

WARNING! When rechargeable 
batteries are installed in the 
CheckPoint, ordinary batteries must 

NOT be installed.

CheckPoint error messages

"P.Er.

"Power error. Error in the external
power supply. Power supply with
incorrect output voltage connected.
Remove the power supply
immediately.
CheckPoint cannot be switched off
before the incorrect power supply
has been removed.

"Er.x "

Internal error. CheckPoint must be
forwarded for service.

Summary of Contents for CheckPoint O2

Page 1: ...PBI 320297 B 02 2014 CheckPoint O2 DE DA SV RU FR ES IT EN PT...

Page 2: ...S COPYRIGHT PBI 320297 B 02 2014 CheckPoint O 2 Distributor Sales Service and Calibration Topac Inc 231 CJC Highway Cohasset MA 02025 Tel 781 7408778 Fax781 740 8779 Email sales topac com internet ww...

Page 3: ...2 CheckPoint O2 PBI 320297 B 02 2014 COPYRIGHT Dansensor A S 7 13 5 14 6 4 1 3 2 7 8 8 9 11 12 10...

Page 4: ...condensing O2 Sensor Electrochemical Range 0 100 Resolution 0 1 Accuracy 0 25 abs 2 rel Temperature compensation 2 rel steady state Response time T95 15 sec Expected sensor lifetime 9 months Recommen...

Page 5: ...10 Brugerguide DA CheckPoint O2 PBI 320297 B 02 2014 COPYRIGHT Dansensor A S NOTER...

Page 6: ...second measuring period is started Reduced measuring time The measuring period can be reduced by pressing while measuring is in progress Memory function CheckPoint remembers up to 10 measurements incl...

Page 7: ...n performing consecutive measurements in areas with large differences in the O2 concentrations the device will reach 95 of the true value during the first measurement 15 sec Example See illustration b...

Page 8: ...During calibration the display alternately shows CAL and 20 9 for 45 seconds CheckPoint has now been calibrated Stop calibration Press to interrupt the calibration process External power supply charg...

Page 9: ...t the needle retainer 9 in the opposite side and press it down until flush with the side plate 6 Test control Switch on the CheckPoint and start measuring with atmospheric air 20 9 Block the gas inlet...

Page 10: ...205 Needle kit 0 5 mm 100 pcs Ser Cpl 280167 Needle kit blunt 10 pcs Ser Cpl 310335 Filter sample gas 0 2 water trap 10 pcs Ser Cpl Spare parts 200720 Pump CheckPoint Ser Cpl 200721 Kit hoses internal...

Page 11: ...16 User Guide EN CheckPoint O2 PBI 320297 B 02 2014 COPYRIGHT Dansensor A S NOTES...

Reviews: