Moving the measuring hose inlet

Any technically gifted user can convert the unit 
from a right-handed into a left-handed model and 
vice versa. 

Note:

 Handle the sensitive internal mechanical 

and electronic components with care when 
converting the unit.
Do as follows to convert the CheckPoint from a 
right-handed model into a left-handed model or 
vice versa:
1. Remove the two side plates (10) and the 

battery cover (7).

2. Remove the needle retainer (8).
3. Switch the gas in and gas out screw joints (9) 

without dismounting the hoses (remember to 
fasten the joints using a tool)

4. Mount the side plates (7) and (10), starting 

with the right-hand plate (be careful not to 
squeeze the hoses)

5. Mount the needle retainer (9) in the opposite 

side, and press it down until flush with the 
side plate

6. Test/control: Switch on the CheckPoint, and 

start measuring with atmospheric air (20.9 
%). Block the gas inlet while measuring (e.g. 
by removing the needle and putting your 
finger on the end of the needle pen). The O

2

 

value must not increase by more than 1 %. If 
this is the case, check that the hoses have 
been mounted correctly in accordance with 
the drawing.

Battery replacement

When the CheckPoint using ordinary batteries 
shows "Lo./BAt.", the batteries must be 
replaced. 
1. Unscrew the screw that secures the battery 

cover (7) on the left side, and dismount the 
cover.

2. Disconnect the connector (13) on the battery 

pack.

3. Take out the battery pack, and replace the 

batteries with 4 new alkaline batteries (type 
AA).

4. Mount the battery pack again (lower/left 

connector).

5. Connect the connector to the battery pack.
6. Mount the cover and screw (be careful not to 

squeeze the wire)

Summary of Contents for CheckPoint O2

Page 1: ...PBI 320297 B 02 2014 CheckPoint O2 DE DA SV RU FR ES IT EN PT...

Page 2: ...S COPYRIGHT PBI 320297 B 02 2014 CheckPoint O 2 Distributor Sales Service and Calibration Topac Inc 231 CJC Highway Cohasset MA 02025 Tel 781 7408778 Fax781 740 8779 Email sales topac com internet ww...

Page 3: ...2 CheckPoint O2 PBI 320297 B 02 2014 COPYRIGHT Dansensor A S 7 13 5 14 6 4 1 3 2 7 8 8 9 11 12 10...

Page 4: ...condensing O2 Sensor Electrochemical Range 0 100 Resolution 0 1 Accuracy 0 25 abs 2 rel Temperature compensation 2 rel steady state Response time T95 15 sec Expected sensor lifetime 9 months Recommen...

Page 5: ...10 Brugerguide DA CheckPoint O2 PBI 320297 B 02 2014 COPYRIGHT Dansensor A S NOTER...

Page 6: ...second measuring period is started Reduced measuring time The measuring period can be reduced by pressing while measuring is in progress Memory function CheckPoint remembers up to 10 measurements incl...

Page 7: ...n performing consecutive measurements in areas with large differences in the O2 concentrations the device will reach 95 of the true value during the first measurement 15 sec Example See illustration b...

Page 8: ...During calibration the display alternately shows CAL and 20 9 for 45 seconds CheckPoint has now been calibrated Stop calibration Press to interrupt the calibration process External power supply charg...

Page 9: ...t the needle retainer 9 in the opposite side and press it down until flush with the side plate 6 Test control Switch on the CheckPoint and start measuring with atmospheric air 20 9 Block the gas inlet...

Page 10: ...205 Needle kit 0 5 mm 100 pcs Ser Cpl 280167 Needle kit blunt 10 pcs Ser Cpl 310335 Filter sample gas 0 2 water trap 10 pcs Ser Cpl Spare parts 200720 Pump CheckPoint Ser Cpl 200721 Kit hoses internal...

Page 11: ...16 User Guide EN CheckPoint O2 PBI 320297 B 02 2014 COPYRIGHT Dansensor A S NOTES...

Reviews: