4. MAINTENANCE, CLEANING AND DISINFECTION
WARNING! Do not use strong chemical cleaning agents such as ‘Dettox’ to sterilise
the light.
Do not use any abrasive materials to clean the light.
CAUTION! The light is not waterproof, so do not use excessive cleaning fluid amounts.
Suitably protect hands.
Wipe the light using clean tissue or soft towelling - pay special attention to the switch
and handles.
Apply to all surfaces a 1% solution of sodium hypochlorite using a soft cloth.
Within 10 minutes, remove the chemical from all surfaces using a damp, soft cloth.
The chemical will cause corrosion of metal parts, if it is not thoroughly removed.
Apply Isopranol to the surfaces – this is an alcohol-based fluid which evaporates without
leaving any residue and can be obtained as an impregnated-cloth applicator.
4.1 Troubleshooting
If the light does not illuminate, change the lamp-bar – see Section 5, below.
If the light still does not illuminate, further investigation should only be undertaken by a
qualified electrician.