The remaining chemical components of the Dario Control Solution are:
•
Buffer 75mM
•
Polyvinylpyrolidone 5%
•
Polyethylene glycol 5%
•
SDS 0.1%
•
Preservative 0.1%
Checking Your Glucose Meter with Dario Control Solutions
To perform a control solution test, you must perform the following three
procedures:
•
Removing the Glucose Meter from the Housing
•
Removing a Test Strip from the Cartridge
•
Performing a Control Solution Test
Removing the Glucose Meter from the Housing
Before performing a control solution test or testing your blood, remove the
glucose meter from the housing. You should always store the glucose meter in
the housing when you are not using it.
When not in use, store the meter in the Dario Housing. Do not
leave your meter connected to your smart mobile device with
the screen in ON mode for more than two (2) minutes, as the
meter can get warm which can affect the accuracy of the
reading.
To remove the meter from the housing:
1.
Hold the housing so that the ridged orange panel is close to you and the
end with the lancing device is facing away from you.
2.
Slide your thumb back on the ridged orange panel in the direction
indicated by the arrow, as shown in the image below. The meter will pop
up.