56
manufactured by Professional Disposables International, Inc. and are available
online through amazon.com and officedepot.com. Before cleaning, remove the
meter from its housing. The meter must not be connected to the mobile
device while cleaning.
Clean the meter, housing, orange lancet cover, and white cartridge cover by
wiping the outside surfaces with a Super Sani-Cloth Germicidal Disposable
Wipe to remove all visible dirt. Remove the white cartridge cover and wipe the
cartridge cap as well. Before you begin, squeeze out any excess liquid from the
wipe so that it is damp but not dripping wet. Take extreme care not to get
liquid on the test strip port connector.
Step 2: Disinfection
Prior to disinfection ensure you have properly cleaned your device as
recommended in
Step 1: Cleaning
. Disinfect the meter, housing, orange lancet
cover, and white cartridge cover using a clean Super Sani-Cloth Germicidal
Disposable Wipe to kill the bloodborne pathogens that may spread infection.
Using a second clean Super Sani-Cloth Germicidal Disposable Wipe:
•
Disinfect the meter on all four (4) surfaces and the audio plug by wiping up
and down at least two (2) times to each direction, as shown in the images
on the following page. Do not allow any moisture to get into the test strip
port connector.
•
Disinfect the housing, orange lancet cover, and white cartridge cover by
wiping the outer surfaces. Ensure that you wipe all parts including the
housing, the white cartridge cover, and the orange lancet cover up and
down at least two (2) times in each direction as demonstrated in the
images on the following page.
•
Remove the white cartridge cover and disinfect the cartridge cap by
wiping up and down thoroughly at least two (2) times in each direction as
demonstrated in the images on the following page.
•
It is very important to let the liquid from the wipe sit on the outer surface
for at least two (2) minutes. Be extremely careful to avoid getting liquid
onto the test strip port connector.
The images below demonstrate how to use Super Sani-Cloth Germicidal
Disposable Wipes to clean and disinfect your Dario System.