67
CHAPTER 7
TECHNICAL INFORMATION AND SPECIFICATIONS
This chapter provides technical information about your Dario System.
How the Glucose Meter Works
The Dario System determines the amount of glucose in fresh capillary blood
using electrochemical biosensor technology. An electrical current is generated
by the reaction of glucose with the reagent on the test strip (glucose oxidase).
The strength of the current produced by the reaction depends on the amount
of glucose in the blood sample. The glucose meter detects the current and
converts it into a blood glucose reading.
Compatibility
Dario is compatible for use on iOS smart mobile devices. For full list of
compatible devices, please refer to
www.mydario.com/support/getting-
started/
Technical Specifications
Meter Specifications
Result Range
20-600 mg/dL (1.1-33.3 mmol/L)
Calibration
Plasma-equivalent
Sample
Fresh capillary whole blood
Test Sample
0.3 microliter
Test Time
6 seconds
Assay Method
Glucose Oxidase Biosensor
Hematocrit Range
20% - 60%
Operating Temperature
50°F-95°F (10°C-35°C)