7
The Dario Control Solutions are for use with the Dario Blood Glucose Meter
and the Dario Blood Glucose Test Strips to check that the meter and test strips
are working together properly and that the test is performing correctly.
System Limitations
The Dario System should not be used for the following purposes:
•
The diagnosis of, or screening for, diabetes.
•
Testing the glucose levels of neonates (children younger than 4
weeks of age).
•
Testing glucose levels of arterial or venous blood.
•
Testing glucose from sites other than the fingertip.
•
Testing patients who are critically ill, in shock, or dehydrated.
If you are taking acetaminophen containing drugs (e.g. Tylenol) or
Vitamin C (ascorbic acid) you may get inaccurate results with this
system. If you have a disease or condition in which uric acid levels
in your blood may be elevated (> 11 mg/dL), such as gout, you may
get inaccurate results with this system. If you have very high levels
of triglyceride (> 810 mg/dL), you may get inaccurate results with
this system. This system should not be used when undergoing
xylose absorption tests.
Important Safety Instructions
The meter and lancing device are for single-patient use
.
Do not share them
with anyone, including other family members. Do not use on multiple patients
.
All parts of the kit are considered bio-hazardous and can potentially transmit
infectious diseases even after you have performed disinfection
.
For important instructions for disinfecting your meter and lancing device, refer
to
CHAPTER 5 CLEANING AND MAINTENANCE
on page
55
.
The U.S. Food and Drug Administration (FDA) and the Centers for Disease
Control and Prevention (CDC) have issued warnings and notifications regarding
the risk of bloodborne pathogen transmission when blood glucose monitoring