background image

01787 370187  

|

  info@darouv.co.uk  

|

  

darouv.co.uk

7

Electrical
 
This unit must be installed by a qualified electrician.

The Saphir range is supplied with a standard, BS1363 
UK mains plug and should be connected to an 
earthed 220-240V, 50-60Hz, single phase and neutral 
supply, via a 3A fused socket.

The control box is provided with a trailing earth 
lead 

– this must be connected to the stud on the 

chamber body

. If the installation of the chamber has 

interrupted a water pipe, bond these together again.

If the UV lamp is switched off at any point wait for 
approximately 30 seconds for the electronics to 
discharge before switching it back on.

Volt Free Contact Connections 
( and Saphir Pro only)

A 4-way (3-ways plus earth) connector is supplied 
to mate to the control box volt-free contact socket. 
If required, a length of suitable three-core plus earth-
control cable should be used to wire between the 
connector and the remote equipment. A junction box 
may be required to make the connection.

No voltages are connected internally to the relay – 
this must be done via the cable plug. The relay drives 
when the lamp is working correctly.

The socket is capped if this feature is not used. 

See specifications, below, for relays ratings.

Connections

Terminal 1 - relay common. Connect to supply.

Terminal 2 - relay normally closed. Supply output 
when lamp has failed.

Terminal 3 - relay normally open. Supply output 
when lamp is OK. Connector Terminal 4 – earth. This 
terminal is bonded to earth inside the control box.

Indicators and Controls

The range has the following features:

•  

Switch to reset the internal hour counter (see 
maintenance, lamp replacement).

•  Three-way indicator display showing the status of 

the system.

•  Alarm to warn of lamp failure and end of lamp life, 

this can be temporarily silenced.

•  Volt-free contact output for remote display of lamp 

on indicator (see specification).

System Status

Power On

Lamp On

Lamp Status

Alarm

System operating correctly,  
lamp less than 11 months old

Green

Blue

Green

Off

System operating correctly,  
lamp between 11 and 12 months old

Green

Blue

Alternating 

Green & Red

Off

Lamp still operating,  
but more than 12 months old

Green

Blue

Red

Bleep on  

12-second interval

Lamp failed

Green

Off

Red

Bleep on  

3-second interval

Saphir Range

Summary of Contents for Saphir Series

Page 1: ...red by Daro UV in the UK other requirements considered Daro UV is a company registered in England Wales Registration Number 06070435 Designed Manufactured in the UK 015 Saphir Range Installation Opera...

Page 2: ...mp failure and end of life Alarm and microcontroller reset button Volt free contact for remote telemetry or solenoid shut off valve connection Selective 254nm UV sensor Digital UV intensity display wi...

Page 3: ...aned and sterilised when the UV unit is fitted To ensure the water does not become stagnant the tank should store no more than a few hours of water supply Safety UV light can harm eyes and skin Never...

Page 4: ...icles down to 5 10 micron in size will normally ensure the water is suitable and such a filter must be fitted unless the water is known to be completely clear at all times Filtration will also extend...

Page 5: ...wall pace apart by about of the length of the chamber with the slots uppermost using the correct screws for your wall type screws not included Open the worm drive clips to their maximum and fit over...

Page 6: ...acket Saphir 10 D Quartz Clamp Key E Saphir 1 2 2 3 0 0 1 Saphir 2 0 2 3 0 0 1 Saphir 3 0 2 3 0 0 1 Saphir 4 0 2 3 0 0 1 Saphir 7 0 2 3 0 0 1 Saphir 10 0 0 0 1 2 1 6 The accessory pack contains the fo...

Page 7: ...to the relay this must be done via the cable plug The relay drives when the lamp is working correctly The socket is capped if this feature is not used See specifications below for relays ratings Conne...

Page 8: ...nged at regular intervals It is not possible to measure the level of UV being produced by the lamp by visual inspection Therefore a regular routine of lamp replacement is recommended to ensure effecti...

Page 9: ...the quartz tube Note Care must be taken when removing quartz tube as it is not supported inside the chamber To avoid breakage it must be kept parallel with the axis of the unit when being changed 4 Cl...

Page 10: ...d steel Painted steel Painted steel Painted steel Painted steel Power supply VAC PH Hz 220 240V 50 60Hz single phase Current rating amps 0 095 0 116 0 163 0 214 0 330 0 330 Fuse external 3A HBC Mains...

Page 11: ...be valid if approved Daro UV Systems spare parts are used All models are to be operated and maintained in accordance with the owner s operating instructions This warranty will become void if the ultr...

Page 12: ...te decreases values Hold the buttons down to scroll through values more quickly To save changes scroll through the entire parameter list to reach the final page configuration write and wait for the se...

Page 13: ...and white buttons to enter the correct code if changed from the default 3 Press the red button again to show freeze This freezes monitoring and disables alarms to allow maintenance This will remain a...

Page 14: ...again to set language Scroll through options with the grey and white buttons 5 Press the red button again to set date and time mm dd yy Change values with the grey and white buttons 6 Press the red bu...

Page 15: ...ower downs of the system Press the white button to clear and the red button to confirm 8 Press the red button again to access startup time This sets a delay from power up to when monitoring starts Ple...

Page 16: ...es Press the white button to clear and red button to confirm 11 Press the red button again to access abs counter This records the total hours the system has run since initiation and is NOT CHANGEABLE...

Page 17: ...NGE 14 Press the red button again to access endvalue This is pre set to 100 Please DO NOT CHANGE 15 Press the red button again to access adjust to calibrate the system Using the up button grey or down...

Page 18: ...ld setting The display flashes red and volt the free contacts will also switch 17 Press the red button again to access delay to add a time delay before the main alarm is triggered 18 Press the red but...

Page 19: ...gain to access delay to add a time delay before the pre alarm is triggered 20 Scroll to config press the red button to confirm The new settings will be applied Settings Setting Default Value Lock code...

Page 20: ...01787 370187 info darouv co uk darouv co uk Notes Saphir Range 20...

Page 21: ...01787 370187 info darouv co uk darouv co uk Notes Saphir Range 21...

Page 22: ......

Page 23: ......

Page 24: ...darouv co uk Unit 9 Ballingdon Hill Ind Est Sudbury Suffolk CO10 2DX Systems designed and manufactured by Daro UV in the UK other requirements considered Daro UV is a company registered in England Wal...

Reviews: