Insufficient Vacuum
If you notice insufficient vacuum on the probe
tips (e.g., devices are dropped), perform the Vac-
uum Check: see
some of the checks fail (red Xs on report when
the
details
check box is checked) then check that:
•
all hose fittings are tight and hoses are not
pinched or clogged.
Perform the probe blowout procedure; see
•
the air filters are not dirty. Replace air filters
if necessary. See
•
the input air to the FLX is at
.48-.55 MPa
(70-80 PSI).
If you can’t find the source of the problem, save
the detailed
Check Vacuum
report and send it to
support@dataio.com.
Failures Picking and Placing
If probes drop devices or fail to pick and place
devices accurately, check the following:
•
check for insufficient vacuum; see previous
heading.
•
check the Z-Offset; see
•
check the module alignment; see
Summary of Contents for FLX500
Page 4: ... Data I O FLX500 ii ...
Page 12: ... Data I O FLX500 Chptr 1 8 blank page ...
Page 28: ... Data I O FLX500 Chptr 2 24 blank page ...
Page 40: ... Data I O FLX500 Chptr 4 36 blank page ...
Page 42: ... Data I O FLX500 Chptr 5 38 Daily Sign off Clean the sockets page 10 Date Initials Notes ...
Page 43: ...5 Sign off Charts Daily Sign off 39 Daily Sign off Date Initials Notes ...
Page 44: ... Data I O FLX500 Chptr 5 40 Daily Sign off Date Initials Notes ...
Page 45: ...5 Sign off Charts Daily Sign off 41 Daily Sign off Date Initials Notes ...
Page 46: ... Data I O FLX500 Chptr 5 42 Daily Sign off Date Initials Notes ...
Page 48: ... Data I O FLX500 Chptr 5 44 Weekly Sign off Date Initials Notes ...
Page 52: ... Data I O FLX500 Chptr 5 48 blank page ...
Page 57: ... FLX500 ...