Maintenance ■ Troubleshooting
- 136 -
Data I/O ■ 096-0465-001C
Programming Problems
Lamp Color
Error Messages
Possible Resolution
None
•
One programmer has
many continuity
failures (in Statistics
window).
•
Move the Socket Adapter to another
programmer to see if the problem is the
programmer or the adapter, as follows:
1.
If a job is running click Pause and wait
until all programmer activity is done.
2.
Check that there are no devices in the
target adapter. If there are, click Finish.
3.
Replace the target Socket Adapter with
another empty adapter.
If the failure moves with the Socket
Adapter, check for bent, damaged or worn
pins on the sockets.
If continuity failures persists in the same
programmer, see
Pick and Place Problems
on page 137.
If a programmer is not performing and you
cannot correct the error, you can continue to use
other programmer sites after disabling the
non-functional programmer. See Disabling
Programmers in the PSV5000
Operators
Manual
or the CH700 on-screen Help.
Red
•
All programmers fail to
program devices.
•
Check that the Programmer Simulation
switch on the Setup window is
OFF
. If not,
switch it to OFF.
•
Check that the Programmers circuit switch
on the Power Panel is On.
•
Check the Log file for helpful information as
follows:
1.
Pause the job.
2.
Exit the Run window.
3.
At the Setup window click System >
Log File.
Summary of Contents for PSV5000
Page 12: ......
Page 36: ......
Page 160: ......
Page 161: ...Electrical Schematics for PSV5000 and Equipment ...
Page 162: ......
Page 163: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 176: ...PROGRAMMABLE MEDIA EXPERTS ...