Shimadzu GC-2014C
4 Using the control module
19
4.5 Shimadzu GC-2014C Setup
Dialog tabs:
Connection
-
Piping
-
Signals
-
Heater Ports
-
Line Config
Shimadzu GC-2014C Setup
dialog (accessible through the
System
Configuration
dialog) allows to view and manually set the
hardware configuration of the chromatograph. All fields
(with the exception of
Name
field and the
Signals
tab) can be
read from the chromatograph by using the
Autodetect
button.
It is highly recommended not to change any data read from
the chromatograph by the
Autodetect
function, because these
settings are checked against those in the GC when opening
the
Instrument
window. If they don't match, the control is not
possible. The only parts of the dialog that can be modified
without risk is the
Name
parameter on the
Connection
tab
and any data on the
Signals
tab.
Connection tab
Fig. 16 Shimadzu GC-2014C Setup - Setup
Main page of the
Shimadzu GC-2014C Setup
dialog.
Port
Selects the COM port to which the
Shimadzu GC-2014C
chromatograph is connected.
Baud Rate
Sets the baud rate of the communication. For a
chromatograph directly connected to the PC via short cable
it is recommended to set the
Baud Rate
to
115200
, for the
chromatograph connected by long serial cable or through