Register Description
Product Reference Guide
19
BIN_SAMPLETIME Registers (RW - 16 words)
Register
Address
Op.
Default
BIN_SAMPLTIME_0
0x23
RW
0x0258
BIN_SAMPLTIME_1
0x24
RW
0x0000
....
.....
0x0000
BIN_SAMPLTIME_15
0x32
RW
0x0000
These registers are used to define the sample interval to apply for the next
acquisitions in the corresponding bin. The value is expressed in seconds.
BIN_THRESHOLD Registers (RW - 16 words)
Register
Address
Op.
Default
BIN_THRESHOLD_0
0x33
RW
0xFFFF
BIN_THRESHOLD_1
0x34
RW
0xFFFF
....
.....
0xFFFF
BIN_THRESHOLD_15
0x42
RW
0xFFFF
These registers represent the threshold value for the alarm (expressed in
seconds) on the corresponding bin. If the bin counter value is greater or
equal than the threshold value the alarm bit for the corresponding bin is set.
Default value is 0xFFFF for each bin that means that alarm is disabled.
MKT_ACTV_ENERGY Registers (RW - 2 words)
Register
Address
Op.
Default
MKT_ACTV_ENERGY_L
0x43
RW
0x0000
MKT_ACTV_ENERGY_H
0x44
RW
0x0000
The value, expressed in floating point 32bits (IEEE-754) notation, represents
the Mean Kinetic activation energy and it is expressed in J/mol. The map
between the floating point fields and the two registers is reported below:
Sign
Exponent
Mantissa
31
30…23 (8 bits)
22…16 15…0 (23 bits)
MKT_ACTV_ENERGY_H sig
MKT_ACTV_ENERGY_H exp
MKT_ACTV_ENERGY_L