Product Reference Guide
5-1
Section 5
Calibration
A number of situations require the scale to be calibrated. They are:
•
at initial installation of the scanner/scale
•
if the scale cannot be re-zeroed
•
if you change the weight measure from pounds to kilograms or
vice-versa
•
if diagnostics indicate a calibration error
•
the weigh module has been replaced
Follow the procedures on the following pages to ensure that the scanner/
scale will meet Weights and Measurement requirements.
Certification of the scanner/scale’s weighing apparatus is subject to
Federal, State and Local Weights and Measures statutes and is
restricted to authorized government agencies and/or duly registered
agents thereof. Anytime a scale is calibrated, it should be properly
sealed with a lead and wire or paper seal before being placed into ser-
vice.
It is your responsibility to check with the appropriate authorities in your
area to ensure compliance with pertinent regulations before removing
any official seals or placing a newly calibrated scanner/scale into ser-
vice.
LEGAL NOTE
Summary of Contents for MAGELLAN 8500?
Page 1: ...Magellan 8500 Product Reference Guide...
Page 14: ...12 Magellan 8500 Scanner NOTES...
Page 66: ...2 36 Magellan 8500 Scanner NOTES...
Page 84: ...3 18 Magellan 8500 Scanner Figure 3 6 Removing Replacing the Gasketed DLC Window a b c d...
Page 89: ...Product Reference Guide 4 5 Figure 4 1 7 Segment LED Display 7 Segment Display Models vary...
Page 98: ...4 14 Magellan 8500 Scanner NOTES...
Page 420: ...D 2 Magellan 8500 Scanner NOTES...
Page 436: ...E 16 Magellan 8500 Scanner NOTES...
Page 452: ...F 16 Magellan 8500 Scanner NOTES...
Page 453: ...Product Reference Guide F 17 NOTES...
Page 454: ...F 18 Magellan 8500 Scanner NOTES...