5-6
Magellan
®
9500
8. The Yellow LED will extinguish for approximately 10 seconds and
the Remote Display will alternately display
and
until the scale is ready to proceed.
9. If the calibration was successful, the speaker sounds a single tone,
the Scale Status LED begins blinking again, and “End-” appears in
the Remote Display.
10. If the calibration was not successful, the speaker will sound five
tones indicating a scale failure, and the Scale Status LED will blink
twice, strobe (fast blinks) and then continually repeat this sequence
until reset. Remove all weight from the Weigh Platter, and repeat
the procedure starting with step
7.
11. Press the Zero Push Button again to permanently store the calibra-
tion data and exit Calibration Mode. You have completed the cali-
bration of the scanner/scale. The unit’s calibration must now be
verified as required by state and/or local weights and measures regu-
lations. The verification procedure must be performed to assure that
a scale will pass Weights and Measures requirements before it is
placed into commercial/retail service.
You have completed the scale calibration procedure. You must now con-
tinue with the calibration verification tests to complete the scale’s calibra-
tion. If you are weighing in pounds, continue with the following
procedures to verify that the scale was successfully calibrated and that it
weighs properly. If you are weighing in kilograms, turn to the instructions
titled,
Calibration Verification (Kilograms)
, later in this chapter.
Summary of Contents for Magellan 9500?
Page 1: ...Magellan 9500 9500 9504 9504 with SmartSentry Product Reference Guide...
Page 28: ...1 16 Magellan 9500 NOTES...
Page 100: ...4 14 Magellan 9500 NOTES...
Page 116: ...5 16 Magellan 9500 NOTES...
Page 396: ...6 280 Magellan 9500 NOTES...
Page 416: ...B 10 Magellan 9500 NOTES...
Page 426: ...D 2 Magellan 9500 NOTES...
Page 442: ...E 16 Magellan 9500 NOTES...
Page 458: ...F 16 Magellan 9500 NOTES...