Adjustments and Maintenance
2-5
Step
Action
Displayed Message
Comment(s)
8
Press the ESC Key to exit the
calibration routine, and then
press the EXIT Key to exit the
Menu System. Finally, to
complete the calibration for die-
cut, notched, and reflective
media, press and hold the
FEED Key until at least one
label has been output.
CALIBRATION COMPLETE
followed by...
READY
The printer is now ready for
use.
2.1.3
Advanced Entry Calibration
ADVANCED ENTRY should only be used when the Standard Calibration method has failed (see Section
2.1.2). To facilitate the placement of the media sensor, the printhead assembly can be raised. During this
procedure, sampled readings will be used to manually establish the best gain algorithm for sensing the
media. The steps below will guide you through the procedure:
Step
Action
Displayed Message
Comment(s)
1
Press the MENU Button.
CANNOT CALIBRATE
-or-
UNCALIBRATED
See the
Operator’s Manual
for
menu navigation tips.
2
Press the ENTER Key to
access MEDIA SETTINGS, and
then use the UP Button to scroll
to SENSOR CALIBRATION.
MEDIA SETTINGS