PAGE 26
UV119UE - DSP 500 : USER MANUAL - v. 4.0
PROBLEM SOLVING
PROBLEM
POSSIBLE CAUSE
REMEDY
After switching on the
External fuse blown
Replace the faulty fuse with new of
same type
scale both displays
remain OFF
Failure of display card
Call for technical service
The scale switches on
but then it blocks and
the message
There is a weight on
scale platter
Take off the weight. Switch off and
then on again.
SEARCH ZERO
WEIGHT
is displayed
Problem with load cell
or with AC/DC
converter
Call for technical service
The scale switches on
but then it blocks and
the message
Platter is not placed on
scale
Place the platter on the scale. Switch
off and then on again
TOO LOW WEIGHT
is
displayed
Problem with load cell
or with AC/DC
converter
Call for technical service
The scale switches on
but then it blocks and
the message
FIND SCALES
is
displayed
Defective connection
with one or more scales
Press key
<C>
and check if the message
SCALE ABSENT
is displayed. Switch
on and/or connect the scales resulting
absent. Press
<PRINT>
The scale cannot detect
the paper
White stamps missing
on paper-cassette
Remove the paper cassette and check
that white stamps are present
Photocell sensor failure
Call for technical service
The scale cannot detect
correctly different paper
cassettes types
White stamps
positioned incorrectly
on the paper-cassette
Remove cassette and check stamps
position. If needed, put stamps
correctly.
Photocell sensor failure Call for technical service
Summary of Contents for DSP 500
Page 1: ...ELECTRONIC SCALE DSP 500 USER MANUAL v 4 0 UV119UE Doc a Ware 960510...
Page 2: ......
Page 4: ...PAGE ii UV119UE DSP 500 USER MANUAL v 4 0...
Page 6: ...PAGE iv UV119UE DSP 500 USER MANUAL v 4 0...
Page 12: ...PAGE 6 UV119UE DSP 500 USER MANUAL v 4 0...
Page 29: ...PAGE 22 UV119UE DSP 500 USER MANUAL v 4 0...
Page 67: ...PAGE 21 UV119UE DSP 500 USER MANUAL v 4 0...
Page 90: ...UV119UE DSP 500 USER MANUAL v 4 0 PAGE 83...
Page 111: ......
Page 112: ...PAGE 104 UV119UE DSP 500 USER MANUAL v 4 0...
Page 113: ...NEW FEATURES PRE SET TARE FUNCTION SELF SERVICE VERSION...