DI-4730 Hardware Manual
Controls, Indicators, and Connections
17
4. Plug the formatted (empty) thumb drive into the device (Drive) while holding the button in.
5. Release the button when the mode LED changes to Red/Black.
6. The device will read and load the configuration files to the thumb drive and revert to Idle mode when complete.
Configuration files include network.json (Ethernet configuration) and config.whd or config.whc (WinDaq
configuration). Use the WinDaq Dashboard to create configuration files.
Apply Ethernet and Stand-alone Configuration to Device
Load Ethernet settings or stand-alone configuration settings using the Control Button. Ethernet and Stand-alone
configurations can also be applied via the Dashboard (preferred method). Device must be in Idle mode for Ethernet or
USB thumb Drive (mode LED blinks green/yellow). The Unlock Code must be purchased for your device in order to
apply the stand-alone configuration.
1. Save configuration files to your USB thumb drive.
2. Confirm device is in Idle mode for USB Flash Drive (mode LED blinks Green/Yellow).
3. Push and hold the Control button.
4. Plug the thumb drive into the device (Drive) while holding the button in.
5. Release the button when the mode LED changes to Red/Black.
6. The device will read and load the configuration files to the device and revert to Idle mode when complete.
Configuration files include network.json (Ethernet configuration) and config.whd or config.whc (WinDaq configura-
tion). Use the WinDaq Dashboard to create configuration files.
Ground Lug
Power Input Jack
Power must be applied with the included power adapter or with an alternate appropriate source (9-36VDC, 15 watts
max). Alternately you may power the device with a Power over Ethernet (PoE) switch.
Power Switch
Power the DI-4730 on/off using the power switch.
Connecting Digital Input Signal Leads
Connect signal leads to the DI-4730 digital inputs:
1. Insert the stripped end of a signal lead into the desired terminal directly under the screw.
2. Tighten the pressure flap by rotating the screw clockwise with a small screwdriver. Make sure that the pressure
flap tightens only against the signal wire and not the wire insulation. Do not over-tighten.
This ground lug should always be connected to a solid chassis or earth ground. This connection is man-
datory when connecting signal leads.