UM-0085-B09
DT80 Range User Manual
Page 394
RG
Upgrading DT80 Firmware
The DT80’s "operating system" (or firmware) is stored in the DT80’s Flash memory. This means that you can easily
upgrade your DT80’s firmware:
•
using a USB memory device, or
•
from a host computer running
DeTransfer
, or
•
using an FTP client
It is strongly recommended that you keep your DT80 up-to-date with the latest firmware. Firmware files are available for
in the Support/ Downloads area.
The same firmware is used for all DT80 series products (DT80/ 81/ 82/ 85). The firmware upgrade file is named
according to its firmware version number – for example,
DT80-9208771.zip
is firmware Version 9.20.8771 (normally
referred to as just "9.20").
Two different firmware packages are available:
•
The standard package (
DT80-9208771.zip
) is suitable for Series 2 or later loggers
•
For Series 1 loggers, use
DT80-9208771-series1.zip
. This contains the same firmware, but the files relating to
the enhanced web interface are not included, as it is not supported on Series 1 loggers.
The above packages are suitable for upgrade using a connected computer running
DeTransfer
.
Packages are also available for upgrade via USB memory device or FTP. These files have a –usbdisk suffix, e.g.
DT80-
9208771-usbdisk.zip
.
Important:
All firmware packages include release notes, which describe the changes made in the release and other information.
Always check the
release notes
for any changes to the upgrade procedure documented here.
Recommended Preparation
In most cases a firmware upgrade can be performed without disturbing the DT80's settings or logged data. If any special
procedures are required, these will be detailed in the firmware release notes.
It is recommended, however, that you carry out the following procedure before upgrading the DT80’s firmware to ensure
that nothing is lost in the event of a problem during the upgrade.
1.
Connect to the DT80 using
dEX
or
DeTransfer
.
2.
Save any previously logged data stored in the DT80’s internal memory by unloading it to the host computer or
copying to a USB memory device.
3.
Using the
dEX
command window, or
DeTransfer
, issue the PROFILE command to return the current profile
settings. Make a note of any non-default settings, which are indicated with an asterisk.
4.
Power the logger from an external supply, and ensure that the internal battery is connected and is fully charged.
5.
Verify that there is sufficient free space on the DT80's internal flash disk, as described below.
You are now ready to perform the firmware upgrade.
Note:
If you have previously used the DT80 web interface then it is recommended that you clear your web browser's cache to ensure
that after the upgrade you are running the correct version of the web interface. (The browser cache, also referred to as "temporary
internet files", holds copies of downloaded files on the computer's hard disk.)
Free Disk Space
As part of the firmware upgrade process, files may be installed on to the internal flash disk (B:). It is therefore necessary
to ensure that there is sufficient free space on this drive before starting the upgrade. To check the free space, type DIR
"B:" in
DeTransfer
, or use the Storage status screen in the enhanced web interface.
The current firmware packages install approximately 13 MB of files (5 MB for Series 1 loggers). If these files do not
already exist then this amount of free space will be required.
When upgrading using a USB memory device or FTP, a temporary copy of the installation package is made on the
internal flash disk. This means that an additional of 15 MB free space (6 MB for Series 1) is required during installation.
The following table summarises the approximate total free space required in order to upgrade to a current firmware
version, based on the previous firmware version in the logger.
Logger Series
PREVIOUS firmware version
Free space required for
DeTransfer
upgrade
Free space required for USB memory
device or FTP upgrade
1
7.02 or later
2 MB
8 MB
1
older versions
6 MB
12 MB
2,3
8.06 or later
5 MB
20 MB
2,3
older versions
10 MB
25 MB