UM-0085-B09
DT80 Range User Manual
Page 37
RG
Category
Channel
Type
Signal\ Sensor Details
Default Channel
Options
Channel
Factor
Output
Units
More Information
Counter
PE
writable
Phase Encoder
Counts
Internal
Maintenance
REFT
Reference temperature of
DT80
or CEM20 terminal
block
scaling
factor
degC
Note 3
Temperature –
Thermocouples (P315)
,
VBAT
Terminal voltage of internal
6V lead acid battery
scaling
factor
V
Battery flat if below 5.6V.
VLITH
Internal Lithium memory-
backup battery voltage.
scaling
factor
V
Replace battery if below
2.8V.
IBAT
Internal main battery current
scaling
factor
mA
Positive if charging,
negative if discharging
VREF
Analog 2.5V voltage source
reference
scaling
factor
V
VZERO
Analog zero voltage reference
scaling
factor
mV
R100
Internal 100 Ohm Shunt
scaling
factor
Ohms
VANA
Internal analog 3.8V rail
voltage
scaling
factor
mV
VDD
Internal 3.3V rail voltage
scaling
factor
mV
VSYS
Internal system supply rail
voltage
scaling
factor
mV
VRELAY
Internal relay supply voltage
scaling
factor
mV
VEXT
Raw voltage onto system
from external supply
scaling
factor
V
CMRR
Common-mode rejection ratio
at maximum gain
dB
Table 3: DT80 Channel Types
Notes
1.
Input termination is on by default (
T
) for DT80/ 81 Series 1 for non-attenuated measurements b and -
terminals
2.
If the current shunt value is specified (as the channel factor) then that value is used. Otherwise, if the
measurement uses the DT80's internal shunt on the # terminal (e.g.
3#I
), then the DT80 uses the actual
calibrated resistance of its shunt. Otherwise, the external shunt is assumed to be 100.0 ohms.
3.
Alternatively, parameter
P36
can be set to force all temperatures to be returned in
degF
,
degR
or
K
.
4.
Offset corrections are subtracted from the measured value.
5.
The bitmask specifies which channels are affected by a multi-bit read or write. Channels where the
corresponding bitmask bit is zero are not affected. For example
1DNO(3)=0
will set digital outputs
1D
and
2D
low but the state of outputs 3D and 4D will be unchanged.
6.
The
delay
channel factor can be used in conjunction with the
R
channel option to generate a fixed width pulse
output.
Note:
use
delay
carefully as it prevents execution of any other schedules, measurements or outputs during the delay.
7.
5Vsw power output will only be active while analog section is powered (Series 3 only) ; use
P21=1
if
continuously power output is required.
Internal Channel Types
The DT80 has its own internal channels, which can be read in exactly the same way as the obvious "external" channels.
Use the channel types below.
Time
The DT80’s real-time clock/calendar has a resolution of 0.1ms, based on a 24-hour clock. Time is read in the same way
as any channel, but without a channel number. That is, sending
T
returns
Time 11:45:10.213