UM-0085-B09
DT80 Range User Manual
Page 47
RG
Category
Channel
Option
M
ut
ua
l
E
xcl
Function
Range of
Option (n)
Comment
Variables
See
/=
n
CV
Divide channel variable by
channel reading.
n
CV
=
n
CV
÷ reading
1 to 1000
time and not during statistical sampling.
=
m
..
n
CV
(
MODBUS
channel only)
Assign values to a range of CVs
Destination
NR
No return
Channels tagged with
NR
are not returned to
the host computer (they may still be logged or
displayed).
NL
No log
Channels tagged with
NL
are not logged
(they may still be returned or displayed).
ND
No display
Channels tagged with
ND
are not displayed
on the LCD (they may still be returned or
logged).
W
Working channel
Same as
(
NR
,
NL
,
ND
)
Working channels are usually used to hold
intermediate values in calculations.
LM
Log alarm measurement
Log the value being tested in an alarm.
Ignored for non-alarm channels.
Output Data
Format
FF
n
default = 1
Fixed-point format
n
=decimal places
0 to 7
Specifies numeric format for display and free
format (
/h
) real-time data.
For example,
FF2
returns
71.46 mV
For example,
FE2
returns
7.14e1 mV
FM
n
uses exponential format if exponent is
less than –4 or greater than
n
Fe
n
Exponential format,
n
=decimal
places
0 to 7
FM
n
Mixed: FF or FE,
n
=significant
digits
0 to 7
BG
min
:
max
Show value as a bargraph on the display.
Channel name
and Units
"
name
"
User-specified name
Default units
ASCII text
Allows channel name and/or units to be
overridden for display and free format (
/h
)
real-time data.
Max 24 characters for user-specified channel
name; 10 characters for units.
"
name
~
unit
"
User-specified name
User-specified units
"
name
~"
User-specified name
No units
"~
unit
"
No channel name
User-specified units
"~"
No channel name
No units
Table 5: DT80 Channel Options