Putting into Operation
R&S
®
NGL200/NGM200
10
Getting Started 1178.8720.02 ─ 06
Risk of radio interference
This instrument is compliant with Class A of CISPR 32. In a residential envi-
ronment, this instrument may cause radio interference.
Risk of instrument damage during operation
An unsuitable operating site or test setup can cause damage to the instru-
ment and the connected devices. Ensure the following operating conditions
before you switch on the instrument:
●
The instrument is dry and shows no sign of condensation
●
The instrument is positioned as described in
●
The ambient temperature does not exceed the range specified in the
data sheet
●
Signal levels at the input connectors are all within the specified ranges
●
Signal outputs are correctly connected and not overloaded
EMI impact on measurement results
Electromagnetic interference (EMI) may affect the measurement results.
To suppress generated electromagnetic interference (EMI):
●
Use suitable shielded cables of high quality. For example, use double-
shielded RF and LAN cables.
●
Always terminate open cable ends.
●
Note the EMC classification in the data sheet.