MPIII Battery Change Quick Start Guide 

1

 

Clean and dry the data logger, making sure 

that the surfaces are clear of contaminants or 
debris. Then, loosen the battery cap with the 
logger gripper provided by holding the logger 
in one hand with the probe facing toward the 
ground, and twisting the battery cap counter-
clockwise with the other hand using the log-
ger cap wrench provided. (PN DT170002)

Mesa Laboratories, Inc.

12100 W. 6th Avenue

Lakewood, CO 80228 USA

Tel: +1-303-987-8000

datatrace.mesalabs.com

datatracesales@mesalabs.com

DataTrace Quick Start Guide

3

 

Remove the old battery by tipping the 

logger to its side until the battery falls out. 
Dispose of the battery appropriately per local 
regulations.

4

 

Examine the battery compartment and 

threads, making sure that the wave springs 
located in the cap and inside the battery 
compartment are in place. If one of the 
springs has fallen out, reinsert it.

Rev. 05OCT2016

Use this guide to help change the battery for a DataTrace MPIII data logger. This quick 
start guide is intended to use the tools provided in the MPIII Maintenance Kit (part num-
ber 124808-500). Replacement batteries part number: 124804-500 (pkg of 20)

2

 

Lift off the battery cap and set it to the 

side, taking care not to misplace the springs 
inside the logger or cap.

5

 

Remove and discard the O-ring to be 

replaced using the pick tool provided. Clean 
the O-ring groove with a clean, lint-free cloth. 
Check for any scratches or nicks in the metal 
before reassembly. If damage is detected or 
contamination is noted, return the logger to 
an authorized servicer for repair.

PN DT170002