DataTrace Quick Start Guide
6
Apply silicone grease to a new O-ring
by rubbing a small amount on your index
finger and thumb and slowly pulling the
O-ring through the grease on your fingers
until the entire O-ring is coated. Then
place the O-ring in the groove of the bat-
tery cap.
7
Holding the logger in the same posi-
tion as step 1, place a new battery in the
battery compartment of the logger, taking
care to put the “+” side of the battery fac-
ing up.
9
Place the logger in the interface. In
DTPro in the Test and Calibration win-
dow, click the “New Battery” button under
the “Tools” menu. Follow the on-screen
prompts to reset the battery and perform
a “test logger.” The new battery is now
successfully installed and the logger is
now ready for use.
8
Install the battery cap in the reverse of
removal. Tighten the cap using the log-
ger cap wrench provided until snug. Wipe
away any excess silicone grease that may
have been pressed out during closure.