22
Installation
DATEQ SPL-D3 manual
UK
Calibration
The calibration allows to test all display LED's and
to calibrate the used microphone.
LED test:
Tests all LED's in green, Orange and Red.
Microphone:
Use a standard approved class 2 calibrator at
94dBA and click on Microphone. The microphone
will now be calibrated to the applied reference source and be internally recalculated to mV/pA.
System
The system page allows to update firmware, backup and restore settings and read the serial
number of the unit.
Language selection and IP settings will be supported in future software and firmware.
Firmware update:
Select latest release firmware and click on update.
The system will check for a valid update.
When a valid firmware is found the display will show
E3 (bootloader mode) on the large display and
update the firmware.
Note;
Some windows builds do not fully support the bootloader mode. When
the progress bar does not start and the display is in E3; disconnect the
USB cable and reconnect it again. The update starts running after
reconnecting.
Settings:
Save settings allows a backup of the current device
settings.
Load settings allows to restore previous saved
settings.
Restore to factory default allows to restore all
settings to factory default. All previous settings will
be lost.
Device:
Shows the serial number of the device, along with
the physical hardware address (MAC address) of
the ethernet connection. These hare hardware
programmed and can not be changed.
In future update the IP address will be added.
Language:
Shows the current software language. Select to change the language of the software.