Aespire View
10-20
M1132382
When the ventilator switches to the backup mode, the alarm text
“Backup Mode active’ displays and remains in the low priority
message site until PSVPro is reinstated or until another mode is
selected. To reactivate the PSVPro mode the user must go into the
Ventilation Mode menu and select PSVPro. Upon selecting PSVPro
the ventilator will immediately begin providing pressure supported
breaths to the patient using the established settings.
PSVPro settings include:
•
Pinsp - sets the target airway pressure.
•
RR - sets the rate of mechanically driven breaths.
•
Tinsp - sets the inspiration time for a mechanical breath.
•
Psupport - sets the pressure support level.
•
Pmax - sets the maximum airway pressure.
•
PEEP - sets the positive end expiratory pressure.
•
Rise Rate - sets the amount of time to attain Pinsp.
•
End of Breath - the drop in inspiratory flow from the peak
inspiratory flow level where the ventilator stops pressure support
mechanical inspiration and begins exhalation.
•
Trigger - sets the flow trigger level.
•
Trig Window - sets the range in percent of the exhalation phase
where a patient may trigger a mechanical breath.
•
Backup mode active - sets the time for backup mode to activate.