5 Preoperative Tests
M1132382
5-7
Vaporizer installation
WARNING
Use only Datex-Ohmeda, Inc. Selectatec series
vaporizers Tec 4 or greater.
w
Do not use a vaporizer if it lifts off the manifold when the
lock lever is in the locked position.
w
Do not use the system if more than one vaporizer can be
turned on at the same time.
w
Tec 6 series vaporizers need the power cable to go
through the channel on the bottom of the vaporizer for
proper alignment. Do not put the power cable on the top
of the manifold or between the vaporizers.
1. Place the vaporizer straight on the manifold.
2. Make sure that the top of each vaporizer is horizontal.
•
If a vaporizer is not horizontal, remove it and reinstall it.
3. Set each vaporizer lock lever to the locked position.
4. Try to lift each vaporizer straight up off the manifold.
•
Do not pull forward on the vaporizer.
•
Do not rotate the vaporizer on the manifold.
5. If the vaporizer lifts off of the manifold, install it again and repeat
this
procedure. If the vaporizer lifts off a
second time, do not use the system.
6. For a Tec 6 series vaporizer:
•
Connect the vaporizer to an electrical outlet.
•
Hold the Silence alarm touch key (minimum of 4 seconds).
•
Make sure all indicators turn on and that the alarm tone
occurs.
•
Release the Silence alarm touch key.
7. Try to turn on more than one vaporizer at the same time.
•
Test each possible vaporizer combination.
•
If more than one vaporizer turns on at the same time, remove
the vaporizers, install them again, and repeat the
procedure. If the test continues to fail, contact a
Datex-Ohmeda trained service representative.
Summary of Contents for Aespire View
Page 1: ...Aespire View User s Reference Manual Software Revision 6 X...
Page 16: ...Aespire View 1 8 M1132382...
Page 46: ...Aespire View 3 16 M1132382...
Page 50: ...Aespire View 4 4 M1132382...
Page 88: ...Aespire View 7 8 M1132382...
Page 112: ...Aespire View 9 10 M1132382...
Page 148: ...Aespire View I 4 M1132382...