Aespire View
5-14
M1132382
Breathing circuit test
1. Set the System switch to On.
2. Set the Bag/Vent switch to Bag.
3. Occlude the bag port.
4. Set the APL valve to 70 cmH2O.
5. Set the O2 flow to 250 ml/min.
6. Occlude the patient connection.
7. Push the O2 flush button and pressurize the bag to approximately
30 cmH2O.
8. Release the O2 flush button. The pressure must not decrease.
Any pressure decrease shown on the pressure gauge indicates a
leak. Repair any leaks in the breathing circuit.
APL valve test
1. Set the System switch to On.
2. Occlude the patient connection.
3. Occlude the bag port.
4. Set the APL valve to 70 cmH2O.
5. Set the O2 flow to 3 l/min. Make sure that the value on the
inspiratory pressure gauge does not exceed 85 cmH2O. Some
pressure fluctuation is normal.
6. Set the APL valve to MIN.
7. Make sure that the value on the inspiratory pressure gauge is
less than approximately 5 cmH2O.
8. Push the O2 flush button. Make sure that the value on the
inspiratory pressure gauge stays near zero.
9. Set the O2 flow to minimum and make sure that the value on the
inspiratory pressure gauge does not decrease below 0 cmH2O.