8 Setup and Connections
M1132382
8-9
WARNING
The filters must be in place to help prevent dust and
particles from entering the breathing circuit.
7. When replacing the canister, make sure that it is seated properly
on the support pins or in the EZchange canister module before
latching it into place.
Electrical connections
Outlets
Labels show outlet voltage ratings and circuit breaker amp ratings.
These are isolated outlets. Regularly test the leakage current.
WARNING
Equipment connected to electrical outlets that are not
isolated outlets can increase the leakage current.
Regularly test the leakage current.
A
B
.82
p0
01
A
B
.7
4p0
49