10 Specifications and Theory of Operation
M1132382
10-13
Suction regulators (optional)
Auxiliary O2 flowmeter (optional)
Venturi Suction Regulator
Performance Category
Pharyngeal Suction
Supply
Air or O2 from system gas supply
Drive Gas Consumption*
28 l/min with pipeline drive gas at 280 kPa (41 psi)
52 l/min with pipeline drive gas at 600 kPa (87 psi)
Maximum Vacuum**
600 mmHg with pipeline drive gas at 280 kPa (41 psi)
550 mmHg with pipeline drive gas at 600 kPa (87 psi)
Maximum Flow*
29 l/min with pipeline drive gas at 280 kPa (41 psi)
32 l/min with pipeline drive gas at 600 kPa (87 psi)
Vacuum Gauge Accuracy
±5% of full scale
*Values are approximate.
**Approximate values are at sea level. Performance is reduced at higher elevations.
Continuous Suction Regulator
Performance Category
Pharyngeal Suction
Supply
External vacuum
Maximum Vacuum*
540 mmHg with external vacuum applied of
540 mmHg and 40 l/min free flow
Maximum Flow*
39 l/min with external vacuum applied of 540 mmHg
and 40 l/min free flow
Vacuum Gauge Accuracy
±5% of full scale
*Values are approximate.
Supply
O2 from system gas supply
Flow rate
0 to 10 l/min
Accuracy
±5% of full scale; not pressure compensated