Electrical safety
WARNING
The system provides connections for items such as
printers, visual displays and hospital information networks
(only connect items that are intended to be part of the
system). When these items (non-medical equipment) are
combined with the system, these precautions must be
followed:
•
Do not place items not approved to IEC 60601-1
closer than 1.5 m to the patient.
•
All items (medical electrical equipment or non-medical
electrical equipment) connected to the system by a
signal input/signal output cable must be supplied from
an AC power source which uses a separating
transformer (in accordance with IEC 60989) or be
provided with an additional protective earth conductor.
•
If a portable multiple socket outlet assembly is used as
an AC power source, it must comply with IEC 60601-1.
The assembly must not be placed on the floor. Using
more than one portable multiple socket outlet
assembly is not recommended. Using an extension
cord is not recommended. After connecting anything
to these outlets, conduct a complete system leakage
current test (according to IEC 60601-1).
•
Do not connect non-medical electrical equipment
directly to the AC outlet at the wall instead of an AC
power source which uses a separating transformer.
Doing so may increase enclosure leakage current
above levels allowed by IEC 60601-1 in normal
conditions and under single-fault conditions. This may
cause an unsafe electrical shock to the patient or
operator.
w
An operator of the medical electrical system must not
touch non-medical electrical equipment and the patient
simultaneously. This may cause an unsafe electrical
shock to the patient.
Summary of Contents for Aespire View
Page 1: ...Aespire View User s Reference Manual Software Revision 6 X...
Page 16: ...Aespire View 1 8 M1132382...
Page 46: ...Aespire View 3 16 M1132382...
Page 50: ...Aespire View 4 4 M1132382...
Page 88: ...Aespire View 7 8 M1132382...
Page 112: ...Aespire View 9 10 M1132382...
Page 148: ...Aespire View I 4 M1132382...