The active low flow system with a 25 mm barb connector is for
use with low vacuum disposal systems. It requires an external
venturi/ejector system with a 40 to 50 l/min extract flow. A flow
indicator on the system indicates when the unit is in operation.
•
The active low flow system with 30-mm ISO taper is for use with
low vacuum disposal systems. It requires an external venturi/
ejector system with a 40 to 50 l/min extract flow. A flow indicator
on the system indicates when the unit is in operation.
See “
” for additional scavenging connection
information.
Connecting active
AGSS with a flow
indicator
To use the optional active AGSS on a system that has a flow
indicator, connect it as follows:
1. Connect the proper hose to the AGSS outlet connector on the
bottom of the AGSS, beneath the breathing system. Attach the
other end to the hospital disposal system.
2. With the AGSS operating, verify that the flow indicator ball on the
flow indicator rises to the green zone, indicating adequate flow.
Note
The ball in the upper red zone indicates excessively high extraction
flow. The ball in the lower red zone indicates extraction flow rate is
too low or the filter is blocked.
3. Complete the tests in the
section of this
manual.