Aestiva
3-24
1006-0938-000
Set SIMV and PSVPro controls
The SIMV and PSVPro modes allow the user to set additional ventilator controls.
The Pinspired, Rate, Tinspired, Psupport and PEEP controls can be set using the
selection keys. The Plimit, Trigger Window, Flow Trigger Level, and Inspiratory
Termination Level may be set through the Setup/Calibration menu.
Step 1
Select the SIMV/PSVPro Setup menu.
• Push the menu key.
• Turn, then push the knob to select
Setup/Calibration.
Step 2
Push the knob to select
SIMV/PSVPro Setup.
w
WARNING
Most anesthetic agents will cause patients to have reduced ventilatory
responses to carbon dioxide and to hypoxemia. Therefore, triggered
modes of ventilation may not produce adequate ventilation.
w
WARNING
The use of neuromuscular blocking agents will reduce the patient’s
breathing response, which will interfere with triggering.
AB.29.0
1
3
AB.90.036
AB.90.03
7
AB.90.045