Aestiva
3-36
1006-0938-000
Connecting Active AGSS
without a flow indicator
The active AGSS option without a flow indicator is for use only with hospital
disposal systems having their own visual indicator of disposal flow rate.
The recommended nominal disposal system flow rate is 36 L/min but will
operate satisfactorily between 30 and 100 L/min.
To use the optional active AGSS installed on the system which does not have a
flow indicator, connect it as follows.
Step
1
Connect a nominal 1/2 inch ID hose to the AGSS
outlet hose connector (a) on the rear of the breathing
system base.
• The hose should be flexible and reinforced to help
prevent kinking and crushing.
Step
2
Attach the other end of the hose to the hospital disposal system.
Step
3
Before you use the system, complete the Preoperative Test procedure. Refer to the Appendix "Preoperative Tests".