Aestiva
A-12
1006-0938-000
Every time a different clinician uses the system
The following tests must be performed every time a different clinician uses the
anesthesia system.
w
W
W
W
WA
A
A
AR
R
R
RN
N
N
NIIIIN
N
N
NG
G
G
G
Do not use a system with a low-pressure leak. Anesthetic gas will
go into the atmosphere, not into the breathing circuit.
Low-pressure leak test
Negative low-pressure leak test
1.
Turn on the auxiliary common gas outlet (AUX some models).
2.
Or, access the common gas outlet (CGO).
3.
Test the leak test device:
AUX
AB.23.063
CGO
AB.23.099
AB.23.063
>60 sec