Aestiva
A-14
1006-0938-000
ISO 5358 or BSI
low-pressure leak test
w
C
C
C
CA
A
A
AU
U
U
UTTTTIIIIO
O
O
ON
N
N
N::::
You can only do a positive pressure test at the common gas outlet.
1.
Access the common gas outlet (CGO).
2.
Connect the leak test device to the common gas outlet with a section of
tubing.
•Turn off the auxiliary common gas outlet.
•Set the Outlet switch to common gas outlet.
•Keep the flow tube vertical for accurate results.
3.
Fully close all flow controls.
4.
Fully open the needle valve on the test device.
CGO
AB.23.099
AB.23.101