Appendix - Preoperative Tests
1006-0938-000
A-19
5.
With a Tec 6 vaporizer:
•Make sure that the vaporizer is under the Tec 6 electrical outlet.
•Make sure that the vaporizer is connected to an electrical outlet.
•Hold down the alarm silence switch for a minimum of four seconds.
•Make sure all indicators come on and the alarm speaker starts.
•Release the alarm silence switch.
•Do not continue until the operational indicator comes on. The
concentration control will not turn if the operational indicator is off.
6.
Try to turn on more than one vaporizer at the same time:
•Test each possible combination.
•If more than one vaporizer turns on at the same time, remove the
vaporizers, install them again, and complete steps 1 through 6.
Breathing system tests
Refer to the applicable operation and maintenance manuals.
At a minimum:
1.
Make sure that the auxiliary equipment (humidifier, etc.) operates
correctly.
2.
Absorbers with active scavenging have a flow tube on the side. Make sure
that it shows a flow in the green (normal) region.
3.
With a circle breathing module, push the drain button for 10 seconds or
more to drain condensate into the absorber.