4b Tests and Calibration 1.X/3.X Software
1006-0453-000 05/04
4b-19
4b.5.4 System Error Log
The
System Error Log
displays the 10 most current errors that have been
logged. The
Error code
, the
Address
, and the
software revision
are important
pieces of information if technical support is required. Note that the system time
is listed in ticks. One tick equals 10 ms.
Software note
The
Bootup Count When Log Last Cleared
entry is not supported in revision
1.X software.
Remarks
The software clears the errors that have been logged when “Clear Error Log” is
set so a new listing may be compiled.
4b.5.5 Alarm log
The software displays the 10 most current errors that have been logged. All
alarms are listed in section5, Troubleshooting.
Note that the system time is listed in ticks. One tick equals 10 ms.
Software 3.X only