5 Troubleshooting
5-4
05/04 1006-0453-000
10035
Non-volatile memory EEPROM data out of
range.
(This is NORMAL if you have just loaded
new software.)
Reboot system. If problem continues, replace CPU board.
10036
Non-volatile memory EEPROM region
protected value error.
(This is NORMAL if you have just loaded
new software.)
Reboot system. If problem continues, replace CPU board.
10037
Software error storing parameter in non-
volatile memory.
Reboot system. If problem continues, replace CPU board.
10038
One of three safety relevant parameter
areas has difference in SRC triplicate
storage (external RAM problem).
Reboot system. If problem continues, replace CPU board.
10039
All three safety relevant parameter areas
are different in SRC storage (external RAM
problem).
Reboot system. If problem continues, replace CPU board.
1003A
There are no more EEPROM areas to switch
to (EEPROM problem).
Reboot system. If problem continues, replace CPU board.
1003B - 1003D
Minimum system monitoring caused by
memory failure (external RAM problem).
Reboot system. If problem continues, replace CPU board.
1003E
Minimum system monitoring caused by
manifold pressure sensor failure.
Reboot system. If problem continues, replace SIB board.
1003F - 10041
Software upset. Sample causes includes
too many key pushes too fast.
Reboot system. If problem continues, replace CPU board.
10042
The system clock has been reprogrammed
(CPU timer problem).
Reboot system. If problem continues, replace CPU board.
10043
Software time-out
Reboot system. If problem continues, replace CPU board.
10044
Memory error (external RAM)
Reboot system. If problem continues, replace CPU board.
10045
Sustained airway pressure holding for
more than 15 seconds.
No Service Action Required. Error only indicates occurrence.
10047
Minimum system shutdown:
SIB 12V test failed
Reboot system. If problem continues,
For machines with original CPU:
1. Measure the voltage at TP9 of the CPU (Section 6.6.2)
Verify SIB_12V = +12V ±5%.
If out of spec, disconnect CPU and measure again.
If still out of spec, replace the CPU; otherwise continue.
2. Replace SIB.
For machines with Integrated CPU:
1. Replace CPU.
2. Replace SIB.
10048
Minimum system shutdown:
15V Local 10 VA test failed
Replace PSB (power supply board).
Replace CPU board
Error code (hex)
Description of error logged
Service action
Summary of Contents for Aestiva 7900 SmartVent
Page 1: ...Aestiva 5 7900 Anesthesia Ventilator Technical Reference Manual ...
Page 44: ...Notes 2 28 05 04 1006 0453 000 ...
Page 82: ...Notes 4a 36 05 04 1006 0453 000 ...
Page 116: ...Notes 4b 34 05 04 1006 0453 000 ...
Page 150: ...Notes 5 34 05 04 1006 0453 000 ...
Page 158: ...Notes 6 8 05 04 1006 0453 000 ...
Page 182: ...Notes 7 24 05 04 1006 0453 000 ...
Page 202: ...Notes 8 20 05 04 1006 0453 000 ...
Page 203: ......