5 Troubleshooting
5-6
05/04 1006-0453-000
1005B
Minimum system shutdown:
+15V analog out-of-range
Refer to section 5.3, Alarm messages.
1005C
Minimum system shutdown:
-15V analog out-of-range
Refer to section 5.3, Alarm messages.
1005D, 1005E
External communications error.
Reboot system. If problem continues:
1. Make sure any external device is correctly connected.
2. Make sure the correct communications protocol is selected.
3. Replace CPU board.
1005F
Unexpected reset in boot code. This can
occur if the system switch is cycled (On-
Off-On) too fast.
Reboot system. If problem continues, replace CPU board.
10060
Stuck in boot code
Reboot system. If problem continues, replace CPU board.
10061
Minimum system monitoring caused by
incorrect feed back on flow valve current.
Reboot system. If problem continues:
1. Check power supply voltages.
2. Replace CPU board.
20001
Possible CPU board A/D converter
problem.
Reboot system. If problem continues, replace CPU board.
20002-2000C
Software upset
Reboot system. If problem continues, replace CPU board.
2000D
A switch for EEPROM region was requested
but there are no more areas to switch to
(EEPROM problem).
Reboot system. If problem continues, replace CPU board.
2000E-2000F
Software upset
Reboot system. If problem continues, replace CPU board.
20010
Software upset
Reboot system. If problem continues:
1. Check
Total System On Time
on the System Error Log Screen.
If unusually large, upgrade software to Rev 3.5 or greater.
2. Replace CPU board.
9999900
Bootup RAM failure
Reboot system. If problem continues, replace CPU board.
9999901-
9999918
CPU upset
Reboot system. If problem continues, replace CPU board.
9999919-
999991F
Software upset
Reboot system. If problem continues, replace CPU board.
9999920
Flash CRC test failed
Reboot system. If problem continues, replace CPU board.
9999921
Software download failed
Try to load the software again. Reboot system. If problem
continues, replace CPU board.
9999922
Stuck in boot code.
Reboot system. If problem continues, replace CPU board.
9999923
Unexpected interrupt or jumped into boot
code unexpectedly. Cycling power too fast
can cause this.
Reboot system. If problem continues, replace CPU board.
10010100-
10010211
Serial port error:
100 = parity
203 = baud rate
204 = not initialized
207 = out of sync
Reboot system, if problem continues:
1. Make sure any external device is correctly connected.
2. Make sure the correct communications protocol is selected.
3. Do the serial ports test (Section 4b.4.7).
4. Replace CPU board.
Error code (hex)
Description of error logged
Service action
Summary of Contents for Aestiva 7900 SmartVent
Page 1: ...Aestiva 5 7900 Anesthesia Ventilator Technical Reference Manual ...
Page 44: ...Notes 2 28 05 04 1006 0453 000 ...
Page 82: ...Notes 4a 36 05 04 1006 0453 000 ...
Page 116: ...Notes 4b 34 05 04 1006 0453 000 ...
Page 150: ...Notes 5 34 05 04 1006 0453 000 ...
Page 158: ...Notes 6 8 05 04 1006 0453 000 ...
Page 182: ...Notes 7 24 05 04 1006 0453 000 ...
Page 202: ...Notes 8 20 05 04 1006 0453 000 ...
Page 203: ......