5 Troubleshooting
1006-0453-000 05/04
5-33
5.6.3 Power supply (Integrated CPU)
Note
The power supply outputs are +16 volts and +12.5 volts.
• The +16 volts is used to charge the battery. It is only available when the
machine is connected to an AC supply.
• The +12.5 volts is used to generate the regulated voltages on the CPU board.
In case of power outage, the CPU power circuits are supplied by the battery.
TP
Name
Typical Value
Description
1
+15V
+15V
15V supply
2
-15V
-15V
Analog -15V supply
200
VCHGR
+16V
From power supply
201
PGND
0V
Power (chassis) ground
202
VMAIN
+12.5V
From power supply
207
AGND
0V
Analog ground
209
+5.8V
+5.8V
Supply for inlet valve, flow valve, speaker
213
VH_EL
+14.5V
Supply for EL display
214
+12VLP
+12V
Supply for light package
J200
TP200
TP202
TP2
TP1
TP207
TP201
J3
LS1
TP209
TP214
J201 J203
J202
J2
J5
J6
J8
U14
U23
12 V (+)
16 V (–)
16 V (+)
12 V (–)
12 V (–)
12 V (+)
12V
Battery
Black
Red
L
N
GND
TP213
Integrated
CPU
Power
Supply
AB.29.090
Summary of Contents for Aestiva 7900 SmartVent
Page 1: ...Aestiva 5 7900 Anesthesia Ventilator Technical Reference Manual ...
Page 44: ...Notes 2 28 05 04 1006 0453 000 ...
Page 82: ...Notes 4a 36 05 04 1006 0453 000 ...
Page 116: ...Notes 4b 34 05 04 1006 0453 000 ...
Page 150: ...Notes 5 34 05 04 1006 0453 000 ...
Page 158: ...Notes 6 8 05 04 1006 0453 000 ...
Page 182: ...Notes 7 24 05 04 1006 0453 000 ...
Page 202: ...Notes 8 20 05 04 1006 0453 000 ...
Page 203: ......