6 Maintenance
1006-0453-000 05/04
6-3
6.1 Supply gas inlet filter
1. Element assembly
2. O-ring, install into filter body prior to assembly
3. Filter cap
Figure 6-1 • Filter body assembly on bottom of ventilator manifold
1. Remove the rear cover of the breathing system.
2. Locate the filter cap at the bottom of the vent engine.
3. Unscrew the filter cap and replace the filter.
4. Check the condition of the o-ring on the filter cap. Replace if cracked or
torn.
5. Reinstall the filter and the filter cap.
6. Perform the Preoperative Checkout Procedure in the Aestiva Anesthesia
System Operation Manual.
w
CAUTION
Cross threading the filter bowl can cause debris to accumulate on the
filter resulting in shorter useful life of the filter.
w
CAUTION
Once you start the threads on the filter body, with the O-ring mounted to
the body, do not back out the filter body unless you take it all the way out
and restart the mounting. If the body is partially unscrewed or backed
out from the manifold seat, the O-ring will slip out of place and result in
an unacceptable leak.
1
2
3