1006-0453-000 05/04
7-1
7 Repair Procedures
In this section
This section covers the repair and replacement procedures for the Aestiva 7900
Ventilator and its related components.
7.10.2 To service the MOPV assembly with the molded housing: . . . . . . . . . . . . . . . . . 7-22
w
WARNING
Post-Service Checkout is required after you complete this section. You must
perform Section 3, “Post-Service Checkout,” after performing any maintenance,
service, or repair. Failure to do so may result in patient injury.
w
WARNING
When servicing the ventilator, extreme care must be taken to avoid introducing
foreign debris, particularly metal chips generated by screw threads, into the
pneumatic flow passages of the ventilator. Failure to do so can result in damage
to the flow valve and possible injury to the patient.
w
WARNING
The ventilator must be off and unplugged before you perform any disassembly
procedures to avoid injury.